Commissioning

PR 5220 Instrument Manual

EN-96

Sartorius

4.7

Configuring General Parameters

The configuration of parameters which are not related to the weighing electronics is divided into several
sections (see Chapter 4.3.8).

4.7.1

Serial Interfaces [Serial ports parameter]

To configure the serial interfaces, select

-[Serial ports parameter].

Setup

Serial ports parameter

Select [Serial ports parameter] with

Operating parameter

and

Setup/

Serial ports

Remote display

Builtin RS485

SMA

none

Param

Config

Remote display protocol [Remote display]
Select the serial interface to which the remote display is connected and then select [Param] to define the
[Baudrate] and the remote display type [Mode] connected.

Setup/Serial ports/

Builtin RS485

Assigned to

Remote display

Protocol

Remote display

Baudrate

9600 bd Select [Baud rate] with

Bits

7

and set the baud rate with

.

Parity

even

Stopbits

1

Mode

multiple transmitters Several remote displays are connected.

Device ID

A Address of the instrument

Next Device ID

B Next address of the instrument

You can only set the baud rate to 300, 600, 1200, 2400, 4800 or 9600. The other parameters (displayed in
light-gray) cannot be changed.
If only 1 instrument is connected to a remote display (normal case), [Mode] must be set to [single transmitter].
If more than 1 instrument is connected to 1 remote display, [Mode] must be set to [multiple transmitters].
At [Device Id] the own instrument address (A, B, C ...) has to be entered, at [Next Device Id] the address of the
subsequent instrument has to be entered

Press

to return to the menu ‚Serial ports’.