PR 5220 Instrument Manual

Commissioning

Sartorius

EN-87

4.6.7

Assigning Load Cells

The load cells (serial number) can be assigned to the place of installation in this menu. This is important for
correcting the dead load (distribution to the individual load cells), for corner correction and in the event of
load cell replacement.

Note:

The assignment from the installation should be documented in the case of load cells being
replaced.

Access the menu as follows:

-[Weighingpoint]

-[Assign]-[View].

Weighingpoint/WP A/

View

LC 1

101

1.958 t

LC 2

103

2.118 t

Load cell load

LC 3

104

1.610 t

LC 4

102

1.828 t

Make sure to unload the scale.

Press [Assign].

Info

Assign

?

WARNING!

Assigning the load cells

deletes the dead load

information!

Cancel

Continue

Press [Continue] to reset the dead load information.
Press [Cancel] for no assigning start.

The load cells are assigned by placing minimum weights (50 kg) on the scale.

Weighingpoint/WP A/

View

Attach calibration weight to area of weighing
construction to be assigned to load cell No 1.
As soon as the device detects the weight
change, the corresponding line is selected.
Confirm the assigning of the 1st load cell

with

.

The future LC no. appears at the far right of the
line.
Remove the weight.
Repeat these steps for load cells 2–4.
Press the [Accept] soft key.

LC 1

101

1.958 t 1

LC 2

103

2.118 t

LC 3

104

1.610 t

LC 4

102

1.828 t

Press

to return.

Info

Accept