Commissioning
PR 5220 Instrument Manual
EN-88
Sartorius
Weighingpoint/
WP A
Press the [View] soft key.
The new assignment will be displayed.
Check the corner load (dead load); see
Chapter 4.6.9.2.
Type
Pendeo process
No of LC
4
LC 1
101
LC 2
103
LC 3
104
LC 4
102
Press
to return.
Search
View
Calib
4.6.8
Calibrating Load Cells
Note:
[Modify] may be used only for minor changes (e.g. changing the dead load); otherwise, always use
[New].
Weighingpoint/
WP A
Type
Pendeo process
No of LC
4
LC 1
101
LC 2
103
LC 3
104
LC 4
102
Press [Calib].
Search
View
Calib
Weighingpoint/WP A/
Calibration
Number of vessel feet
4
Local gravity
9.81379 m/s
2
Max
10000 d
200.000 t
Max: the sum of the max. load cell capacity is
preset (4x 50 t = 200 t).
Scale interval
1 d
0.020 t
Dead load
CAL weight
Press [New].
New
Modify
Param
When you press [New], the data is set to default first and calibration is started.