Commissioning

PR 5220 Instrument Manual

EN-88

Sartorius

Weighingpoint/

WP A

Press the [View] soft key.
The new assignment will be displayed.
Check the corner load (dead load); see
Chapter 4.6.9.2.

Type

Pendeo process

No of LC

4

LC 1

101

LC 2

103

LC 3

104

LC 4

102

Press

to return.

Search

View

Calib

4.6.8

Calibrating Load Cells

Note:

[Modify] may be used only for minor changes (e.g. changing the dead load); otherwise, always use
[New].

Weighingpoint/

WP A

Type

Pendeo process

No of LC

4

LC 1

101

LC 2

103

LC 3

104

LC 4

102

Press [Calib].

Search

View

Calib

Weighingpoint/WP A/

Calibration

Number of vessel feet

4

Local gravity

9.81379 m/s

2

Max

10000 d

200.000 t

Max: the sum of the max. load cell capacity is

preset (4x 50 t = 200 t).

Scale interval

1 d

0.020 t

Dead load

CAL weight

Press [New].

New

Modify

Param

When you press [New], the data is set to default first and calibration is started.