Pr 5220 instrument manual j-bus/modbus protocol – Rice Lake PR5220 Ethernet Transmitter User Manual
Page 121
PR 5220 Instrument Manual
J-Bus/ModBus Protocol
Sartorius
EN-121
Function 5: Writing a Bit
Command
Device
address
Function
number
Address of
the bit
Value of
the bit
Always 0
1 byte
1 byte
2 bytes
1 byte
1 byte
Range
0...127
5
0...127
0 or 255
0
Reply
Device
address
Function
number
Address of
the bit
Value of
the bit
Always 0
1 byte
1 byte
2 bytes
1 byte
1 byte
If the address of the bit is out of the permissible range (0...127), an error message is sent as a reply.
Example of function 5 for setting bit 113 (taring) with ModBus-TCP:
Command
47
13
0
0
0
6
0
5
0
113
255
0
Reply
47
13
0
0
0
6
0
5
0
113
255
The individual bytes are shown.
Function 6: Writing a Word
Command
Device address Function
number
Word address Value of the
word
1 byte
1 byte
2 bytes
2 bytes
Range
0...127
6
0...63
Reply
Device address Function
number
Word address Value of the
word
1 byte
1 byte
2 bytes
2 bytes
If the address is out of the permissible range (0...63), an error message is sent as a reply.