PR 5220 Instrument Manual
Commissioning
Sartorius
EN-111
4.10.1
Adapting the Analog Output
The output current can be adapted in small ranges. This is required, if small deviations from the nominal value
occur in a connected PLC.
Open the menu with
[Show HW-slots]:
Info/
HW-Slots
Builtin
RS485
Builtin
analog out
Builtin
digital i/o
Select [analog i/o].
Info/
HW-Slots
Builtin
analog out
In use by PLC task
1
Analog output
4.004 mA
counts
10934 cnt
Stop PLC Stop I/O
Adjust
Reset
Press [Adjust].
Adapt analog output ?
Safety prompt: Reply [Yes] to start adapting.
Info/HW-Slots/
Adjust Analog Output
Output
4.000 mA
Measured
4.004 mA
Enter e.g. the value for 4 mA measured by the connected PLC under [Measured].
After pressing , the 2nd value (20 mA) is displayed:
Info/HW-Slots/
Adjust Analog Output
Output
20.000 mA
Measured
20.010 mA
Enter e.g. the value for 20 mA measured by the connected PLC under [Measured].