Pr 5220 instrument manual fieldbus interface – Rice Lake PR5220 Ethernet Transmitter User Manual
Page 143
PR 5220 Instrument Manual
Fieldbus Interface
Sartorius
EN-143
Register 5: Device Type and Software Release
Only reading is permitted.
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Byte 0
TYPE MSB
Byte 1
TYPE LSB
Byte 2
MAINVERSION
Byte 3
SUBVERSION
E.g. 5220 Rel 1.23 = 52200123
hex
Register 6: Board Number
Only reading is permitted.
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Byte 0
Board number MSB
Byte 1
""
Byte 2
""
Byte 3
Board number LSB
E.g. 148388723 = 08D83B73
hex
Register 7: (Reserved)
Register 8 ...15: Weight Data
Only reading is permitted.
The gross, net and tare weight are stored as DINT fixpoint.
The real data value is derived from DINT and EXPO as follows :
Value
Real
=reading
DINT
*10
(-EXPO
Register 8
Current gross value
Register 9
Current net value, if tared; otherwise gross
Register 10
Current tare value, if tared, otherwise 0
Register 11
Value on the front-panel display
Register 12
Reserved
Register 13
Reserved
Register 14
Max (FSD)
Register 15
Reserved (free)
Register 24 ... 29: Limit Values (Read/Write)
Register 24
Limit 1 on
Register 25
Limit 1 off
Register 26
Limit 2 on
Register 27
Limit 2 off
Register 28
Limit 3 on
Register 29
Limit 3 off