6 xbpi setting dead load, Xbpi setting dead load, Pr 5220 instrument manual commissioning – Rice Lake PR5220 Ethernet Transmitter User Manual
Page 77
PR 5220 Instrument Manual
Commissioning
Sartorius
EN-77
4.5.6
xBPI Setting Dead Load
Note:
Both terms 'dead load' and 'preload' are used by Sartorius.
Weighingpoint
Weighingpoint A
xBPI-Scale Select [xBPI-Scale] with
and
.
Select [Setup].
Setup
Config
Param
Read the parameters from the xBPI scale with [Setup].
Weighingpoint/
xBPI-Scale Setup
Reading parameters
model
metrologie
Ticks indicate the progress.
device info
settings
An error message displays, unless communication with the xBPI scale is possible!
Weighingpoint/
xBPI-Scale Setup
Calibration
Open with .
Configuration
Select group of specification
Show device info
Weighingpoint/
xBPI-Scale Setup
Dead load
Set
Set dead load.
Delete
Delete dead load.
Span
Adjust with user weight
Adjust with auto weight
Leave with
.
For setting the dead load, remove the weight from the scale and select [Set]. After sending the command, 0 is
indicated on the gross weight display.
Alternatively, the stored dead load can be deleted:
Remove the weight from the scale and select [Delete]. The stored dead load is deleted. The instantaneous dead
load is shown on the weight display.