Commissioning

Sartorius

EN-77

4.5.6

xBPI Setting Dead Load

Note:

Both terms 'dead load' and 'preload' are used by Sartorius.

Weighingpoint

Weighingpoint A

xBPI-Scale Select [xBPI-Scale] with

and

.

Select [Setup].

Setup

Config

Param

Read the parameters from the xBPI scale with [Setup].

Weighingpoint/

xBPI-Scale Setup

Reading parameters

model

metrologie

Ticks indicate the progress.

device info

settings

An error message displays, unless communication with the xBPI scale is possible!

Weighingpoint/

xBPI-Scale Setup

Calibration

Open with .

Configuration

Select group of specification

Show device info

Weighingpoint/

xBPI-Scale Setup

Dead load

Set

Set dead load.

Delete

Delete dead load.

Span

Adjust with user weight

Adjust with auto weight

Leave with

.

For setting the dead load, remove the weight from the scale and select [Set]. After sending the command, 0 is
indicated on the gross weight display.
Alternatively, the stored dead load can be deleted:
Remove the weight from the scale and select [Delete]. The stored dead load is deleted. The instantaneous dead
load is shown on the weight display.