Commissioning

PR 5220 Instrument Manual

EN-48

Sartorius

4.3.6

INFO Function

When you press , the program releases and status messages are displayed. The key also has other
functions; see Chapters 4.4.1.1 and 4.4.9.

Info

Show version

Show status

Show HW-slots

When you select [Show version], the installed program releases and the board number are displayed:

Info/

Version

Firmware

Rel. 02.00.00.00000 Firmware release and creation date

2010-11-11 11:11

PR5220-Application

Rel. 01.20.06 Application release and creation date

2010-11-11 11:11

Bios

Rel 02.00.00.00000 BIOS release and creation date

2010-11-11 11:11

Boardnumber

275401089 Main board identification number

(different from the device serial number)

When you select [Show status], instrument status information is displayed:

Info/

Status

Free system RAM

6328 of 15212 kb

Cal-Switch

closed (‘opened’ if CAL switch is open)

When you select [Show HW-slots], the installed plug-in cards are displayed:

Info/

HW-Slots

Builtin

RS 485 Standard serial interface

Builtin

analog out Standard interface, analog outputs

Builtin

digital i/o Standard interface, digital I/Os

4.3.7

Setup Function (VNC)

Press

to configure the main operating parameters. The instrument configuration depends on the

application and the plug-in cards installed.
Calibration is in a simple dialogue. Compliance with the relevant (verification) standards must be checked by
the person commissioning the instrument or the verification officer. To protect the calibration data from
overwriting, close the CAL switch on the back panel of the instrument. On legal-for trade instruments, the
CAL switch must be sealed in the closed (write-protected) position; see Chapter 4.1.1.