PR 5220 Instrument Manual

Fieldbus Interface

Sartorius

EN-137

Output Area
Data transmission from the PR 5220 to the master is via the output area.
The weight or data information requested by the master is transmitted to the master by PR 5220.
PR 5220 has write access, the master has read access.

Byte Name

Description

0

Read_Value (MSB)

e.g. gross value

1

:::

"

2

:::

"

3

Read_Value (LSB)

"

4

Read_Value _Selected

e.g. gross

5

Write

Active

Power

Fail

Out 3

Out 2

Out 1

Limit 3

Limit 2

Limit 1

Status

6

Cmd

Busy

Cmd

Error

Inp.3

Inp.2

Inp.1

Tare

Active

Cal

Active

Test

Active

Command status

7

Out

Stand-

still

Inside

ZSR

Center

zero

Below

Zero

Over-

load

Above

Max

Adc

Error

Device status

Bit 7

Bit 6

Bit 5

Bit 4

Bit 3

Bit 2

Bit 1

Bit 0

Variable

Function

Read_Value

The weight value is transmitted as a binary 32-bit value with plus or minus sign.
Data type:

DINT

Read_Value_Selected

Acknowledgement of the transmitted value.

Write_Active

The function selected with

Write_Value_Select

is executed once.

This bit is deleted, when

Write_Value_Select

is set to 0.

Power_Fail

Is set when switching on the instrument.
Is reset by ResPower with transition from 01.

Cmd_Busy

The instrument is busy executing a command (e.g. the instrument has received a
taring command and waits for stability).

Cmd_Error

The instrument has interrupted the execution of a command (e.g.

StandStill

could

not be reached within the defined standstill time). The error number can be read at
Lasterror. It is set only, if an action is executed.

Tare_Active

The instrument was tared.

Cal_Active

The instrument is / was configured. When this bit is 1, the scale parameters
(Expo/Unit/Step) must be read again. Set after power On and reset after reading
the max. capacity.

Test_Active

The instrument executes the ADC test. The read weight value is not the gross value,
but the test value.

Out

Above Max. capacity or below zero

Standstill

The instrument is stable.

InsideZSR

The weight value is within the zero setting range.

CenterZero

The weight value is within

center zero

(

0 +/- 0.25 d

)

BelowZero

The weight value is negative (gross

< -0.25d

)

Overload

The weight value has exceeded the Max. capacity, but is still within Max +
permissible overload (gross

<= max. capacity+overload

)

AboveMax

The weight value has exceeded the Max. capacity, but is still within Max +
permissible overload (gross <= max. capacity+overload)

AdcError

A/D conversion error. (Details are given in register 1,

Read_Value_Select = 1

)