Pr 5220 instrument manual commissioning – Rice Lake PR5220 Ethernet Transmitter User Manual
Page 81
Advertising
PR 5220 Instrument Manual
Commissioning
Sartorius
EN-81
Following window appears:
Weighingpoint/
xBPI-Scale Setup
Calibration status
Load to small
Cal-Target
-5000.0 g
--
0.00 g
The following window is displayed after applying the weight:
Weighingpoint/
xBPI-Scale Setup
Calibration status
Difference display Calibration progress without weight.
Cal-Delta
-0.0 g
--
4999.98 g The weight is displayed in high-resolution (10x).
Select [Accept].
Accept
ResError
Abort
The data are saved and the instrument generates a corresponding message:
Weighingpoint/
xBPI-Scale Setup
Calibration status
complete
Net=Grs
5000.0 g
--
5000.00 g The weight is displayed in high-resolution (10x).
Leave with
.
Accept
ResError
Abort