Commissioning

PR 5220 Instrument Manual

EN-38

Sartorius

4.1.2

Factory Settings

Calibration data <default>

Calibration data <default>

Full scale (Max) <3000> <Kg>

Measure time (M) <320>ms

Scale interval <1>

Measuring rate <160>ms

Dead load <0.000000>mV/V

Standstill time <1>M

Span <1.000000>mV/V

Standstill range <1.00>d

Tare timeout <8>M

Calibration parameters <default>

<Absolute> test mode

Overload (range above Max) <9>d

Zero-setting range <50.00>d

* W

& M mode <off>

Zero-tracking range <0.25>d

Filter <off>

Zero-tracking step <0.25>d

Frequency <1.56 Hz>

Zerotrack repeat <0>M

* Parameter W

&M must be set to 'on' or 'off' prior to input of the calibration data, see Chapter 4.4.13.1.

.

4.2

Switching on the Instrument

The instrument can be put into operation and calibrated using a notebook/PC with the VNC program (on the
CD packed with the instrument) and an Internet Browser.

4.3

Configuration and Calibration

There are following possibilities:
-

with VNC viewer (on the enclosed CD-ROM), see Chapter 4.3.4

-

with an Internet Browser (‚Microsoft InternetExplorer’ or ‚Mozilla Firefox Webbrowser’), see Chapter 4.3.5.

The prerequisite is an installed and activated Java (Sun) ’applet’.

4.3.1

Connecting the Device to the Network and Finding out the IP address

The DHCP server is active in the network
An IP address is assigned to the device automatically.

The DHCP server is not active in the network
If the device is connected to a notebook/PC via a point-to-point connection, an IP address is negotiated via
function ‘AutoIP’.

This can take up to 2 minutes!

IndikatorBowser
The IP address can be found out using the ‘IndicatorBrowser’ (supplied on CD-ROM) and via the ‘host name’ of
the device (see also Chapter 4.3.3). The ‘host name’ is composed of the device name and the last 3 bytes of the
MAC ID. A label with the complete MAC ID is fitted inside the door of the instrument.

Hostname: PR5220-6B6A5E