Commissioning, Pr 5220 instrument manual – Rice Lake PR5220 Ethernet Transmitter User Manual

Page 90

Advertising
background image

Commissioning

PR 5220 Instrument Manual

EN-90

Sartorius

[Dead load]
Empty weight as dead load
To use the empty scale/hopper as dead load (normal case):
• Discharge the scale/vessel.
• Press [by load].
The device must now display 0.000.

Weighingpoint/WP A/

Calibration

Number of vessel feet

4

Local gravity

9.81379 m/s

2

Max

10000 d

200.000 t

Scale interval

1 d

0.020 t

Dead load

CAL weight

by load

by value

Known weight value as dead load
To use the known weight value as dead load:
• Press [by value].

An input window appears:

Enter input weight deadload

7.511 t

Ok

Cancel

• Enter the known weight value and press [Ok].
The device must now display 0.000.

[CAL weight]

Press

and an input window appears.

Place CAL weight

on the scale and

enter value

10.000 t

Ok

Cancel

Center the calibration weight on the scale and enter the weight value. Press the [OK] soft key.
Remove the calibration weight.

[Corner correction]

Carry out a corner correction if necessary, see Chapter 4.6.9.

Increased Resolution (10-fold)
During calibration press

to display the weight with 10-fold resolution. The resolution is reset after approx.

5 seconds. Press

if you want to switch to normal resolution immediately.