SMA Protocol

PR 5220 Instrument Manual

EN-132

Sartorius

Reply with ‘A’ and ‘B’ Command
Reply format with ‘A' and 'B' commands (variable length):
<LF><xxx>:<yyyyyy><CR>
The reply format and meaning are:

<LF>

Start of reply from 'A'/'B' command

<xxx>

The field name is three characters long, left-adjusted and filled up with spaces on the
right, if necessary.

The following fields are sent:

“SMA” level/revision

(reply from ‘A’ command)

“MFG” manufacturer marking

(reply from the 1st ‘B’ command)

“MOD” product / model identification

(reply from the 2nd ‘B’ command)

“REV” software version

(reply from the 3rd ‘B’ command)

“SN_“ serial number

(reply from the 4th ‘B’ command)

“END”: This is always the last inscription field

(reply from the last ‘B’ command)

‘:’

Separator between field name and field content.

<yyyyyy>

The data field has up to 25 characters.

The SMA field contains <level/revision>

with the following meaning: level= (1, 2, etc.); revision= (1.0, 1.1, etc.)

<CR>

End of reply from 'A'/'B' command


Example:

Command

Reply

<LF> A <CR>

<LF>SMA:1/1.0 <CR>

<LF> B <CR>

<LF>MFG:Sartorius <CR>

<LF> B <CR>

<LF>MOD:PR 5220 <CR>

<LF> B <CR>

<LF>REV:01.01.9 <CR>

<LF> B <CR>

<LF>SN_:148388723 <CR>

<LF> B <CR>

<LF>END: <CR>


Note: If the controlling computer sends another 'B' command:
<LF> B <CR> <LF> ? <CR>