Installing the Instrument

Sartorius

EN-35

3.1.12

ProfiNet I/O Interface (PR 5220/06 only)

It is a complete ProfiNet I/O interface (SLAVE). It contains a powerful UDP/IP connecting circuitry with transfer
rates of 10 and 100 Mbits/s.

Connection

RJ-45 connecting socket

Transfer rate

10 Mbit/sec and 100 Mbit/sec
Autodetection (10/100, HalfDX/FullDX)

Connection mode

Network

Protocol

ProfiNet I/O

Configuration

XML file (‚GSDML-xxx-Sartorius-PR5220-
xxx.xml’ stored on the CD in directory
‘Fieldbus’)

Certificate

ProfiBus Nutzerorganisation e.V.
for HMS Industrial Networks AB
Certificate no.: Z10006
Report: PN005-1, 12.02.2007

Cable

Twisted pairs, screened,
e.g. patch cable CAT5
Autolink (straight oder crossover

Cable impedance

150 ¥

Potential isolation Yes


Note:

Recommendation for e.g. Siemens S7
Fieldbus slave setting:
Use DHCP [on] as default and activate the master as a DHCP server
(W [assign IP Adr via IO controller]).


Slave – master device names

A unique device name must be assigned out of the master. This name is given highest priority
when establishing the communication.
When changing instruments/servicing, please note:
Apart from the IP address, the device name must correspond to the one of the replacement
device. Explicit assignment out of the master is required.


Example: