4 sma reply messages, Sma reply messages, Pr 5220 instrument manual sma protocol – Rice Lake PR5220 Ethernet Transmitter User Manual
Page 129
PR 5220 Instrument Manual
SMA Protocol
Sartorius
EN-129
Scale Data
Scale Data – First Line
Command: <LF>A<CR>
Reply:
The scale sends the first line of its scale data.
<LF><SMA>:<yyyyyy><CR>
For details, see page 130.
Scale Data – Other Lines
Command: <LF>B<CR>
Reply:
The scale sends additional lines of its scale data.
<LF><MFG>:<yyyyyy><CR>
For details, see page 130.
Scale Information
Scale Information - First Line
Command: <LF>I<CR>
Reply:
The scale sends the first line of its scale information.
<LF><SMA>:<yyyyyy><CR>
For details, see page 130.
Scale Information – Other Lines
Command: <LF>N<CR>
Reply:
The scale sends additional lines of its scale information.
<LF><TYP>:<yyyyyy><CR>
For details, see page 130.
Escape Command
Command: <ESC>
Reply:
This is the only command which does not work according to the <LF>c<CR> protocol principle. It
does not have a reply. The <ESC> character is detected at any time and cancels any current
command.
6.4
SMA Reply Messages
In this section, the replies are described in detail. The data format of each reply has a fixed length. The
communication error is the only exception from this pre-definable format. Thus the controlling computer can
check each reply according to fixed rules, because each data field is in a fixed position.