PR 5220 Instrument Manual

Commissioning

Sartorius

EN-83

4.6

Calibrating Digital Load Cells Type ‘Pendeo®’

4.6.1

General Information

The digital load cells have been calibrated at the factory based on the acceleration of gravity Hamburg,
Germany: 9.81379 m/s². The calibration data in the load cells are invariable. The calibration data for the
gravity acceleration at the place of installation can be adapted only in the instrument and protected against
overwriting (see Chapter 4.1.1). With applications for use in legal metrology, the legal requirements and the
conditions given on the test/approval certificate must be taken into account when selecting the settings.
For connecting digital load cells (xBPI load cells) a Firmware from 2.10 onwards is required.

4.6.2

Viewing the Interfaces

The existing interfaces can be viewed under

-[Show HW-slots]:

Info/HW-Slots

Builtin

RS485 Built-in interface

Builtin

analog out

Built-in interface

Builtin

digital I/O

Slot 4

PR5220/01

Profibus-DP

4.6.3

Selecting and Setting up the Interface

Select

-[Serial ports parameter]-[xBPI-Port] to choose a serial interface.

Setup/

Serial ports

Remote display

-none-

SMA

-none-

xBPI-Port

Builtin RS485 Press

to select RS-485.

Press [Param].

Param

The menu appears.

Setup/Serial ports/

Slot2 RS485

Assigned to

xBPI-Port

Protocol

xBPI

Baudrate

19200 bd

Press

to select the value ‚19200 bd‘.

Bits

8

Parity

odd

Stopbits

1 Press

to select the value ‚1‘.