2 operating parameters, Operating parameters, Commissioning – Rice Lake PR5220 Ethernet Transmitter User Manual

Page 98: Pr 5220 instrument manual

Advertising
background image

Commissioning

PR 5220 Instrument Manual

EN-98

Sartorius

4.7.2

Operating Parameters

Define the basic operating parameters under

-[Operating parameter].

Setup/Operating parameter

AccessCode

0

SetTareKey

tare

& reset tare

SetZeroKey

only when not tared

[AccessCode]
The access code can be used to protect the [Setup] from unauthorized operation. Enter a number with up to 6
digits. As long as you are in this menu, the value can be overwritten as required. When [AccessCode] is set to 0,
no access code ist prompted.

[SetTareKey]
The SetTare function can be configured (VNC operation / Internet browser).
[tare

& reset tare]: the scale is tared if it was not tared previously; otherwise, tare is reset.

[tare

& tare again]: when you press the [Tare] key, the current value is stored in the tare memory and a net

weight of 0 is displayed.
[disabled]: The tare key has no function.

[SetZeroKey]
The SetZero function (VNC operation / Internet browser) can be limited to gross mode with [only when not
tared] or switch the scale to gross mode automatically with [reset tare on zeroset]. If the zero-setting key with
these settings has no effect, the configured zero-setting range (around the zero-point set with the dead load)
is already utilized due to a previous zero-setting operation and/or automatic zero setting.
[disabled]: The key has no function.

Closing the menu
To close the menu, press

. The following message is displayed:

Press [Yes] to save the data.
Press [No] to close the menu without changing data.