PR 5220 Instrument Manual

Installing the Instrument

Sartorius

EN-27

3.1.5

Optocoupler Outputs

The 3 optocoupler outputs have one common potential (GND) for the output group that is separated from the
common potential of the input group.

Connection method

4-pin plug-in terminal block

Cable

Shielded, max. 50 m

Number of outputs

3

Output signal

External supply required

Output current

Max. 30 mA

Output voltage

Max. switching voltage: 28 VDC

Electrical isolation

Yes; a common minus potential for the group
of 3 outputs

Example: relay control connection

When output 1 is active (true), the relay switches.
For protection of the output circuit, relays with
free-wheel diode must be provided.

Example: voltage output connection

When output 1 is active (true), the output voltage
changes from 24/12 V DC into <3 V DC.
A load resistance of 2.2/1 k¥ must be provided.