Page 85: Search for load cells, Pr 5220 instrument manual commissioning
PR 5220 Instrument Manual
Commissioning
Sartorius
EN-85
4.6.5
Adjustment Sequence
During calibration, no data is changed in the digital load cells. The calibration data and parameters are saved
in the instrument. The unique serial numbers of the connected load cells are monitored.
For the adjustment is the following order specified:
• Search for load cells, see Chapter 4.6.6.
• Assign load cells, see Chapter 4.6.7.
• Calibrate ‘New’: Max with weight unit, scale interval, dead load, calibration weight, see Chapter 4.6.8.
Note:
For further information about calibrating weighing points, see Chapter 4.4.
4.6.6
Search for Load Cells
Weighingpoint
Weighingpoint A
Pendeo Load Cells
Press [Assign].
Setup
Config
Param
Assign
Weighingpoint/
WP A
Type
Pendeo process
No of LC
1
LC 1
0
Press [Search].
Search
View
Calib
?
Search for connected load cells.
The actual settings are reset to
default.
Cancel
Continue
Press [Cancel] to accept and display the existing values.
Press [Continue] to start a new search process.