PR 5220 Instrument Manual

Commissioning

Sartorius

EN-85

4.6.5

Adjustment Sequence

During calibration, no data is changed in the digital load cells. The calibration data and parameters are saved
in the instrument. The unique serial numbers of the connected load cells are monitored.
For the adjustment is the following order specified:
• Search for load cells, see Chapter 4.6.6.
• Assign load cells, see Chapter 4.6.7.
• Calibrate ‘New’: Max with weight unit, scale interval, dead load, calibration weight, see Chapter 4.6.8.
Note:

For further information about calibrating weighing points, see Chapter 4.4.

4.6.6

Search for Load Cells

Weighingpoint

Weighingpoint A

Pendeo Load Cells

Press [Assign].

Setup

Config

Param

Assign

Weighingpoint/

WP A

Type

Pendeo process

No of LC

1

LC 1

0

Press [Search].

Search

View

Calib

?

Search for connected load cells.

The actual settings are reset to

default.

Cancel

Continue

Press [Cancel] to accept and display the existing values.
Press [Continue] to start a new search process.