3 fieldbus parameters, Fieldbus parameters, Er 4.7.3 – Rice Lake PR5220 Ethernet Transmitter User Manual
Page 99: Pr 5220 instrument manual commissioning
PR 5220 Instrument Manual
Commissioning
Sartorius
EN-99
4.7.3
Fieldbus Parameters
Setup
Serial ports parameter
Operating parameter
Fieldbus parameter
Press
and
to select [Fieldbus parameter]
This menu item can only be selected if the instrument is a PR 5220/01, a PR 5220/04, PR 5220/06 or
PR 5220/07 version.
The instrument version automatically determines the corresponding protocol:
-
[Profibus-DP] for PR 5220/01
-
[DeviceNet] for PR 5220/04
-
[ProfiNet I/O] for PR 5220/06
-
[EtherNet-IP] for PR 5220/07
Dependent on the interface type, additional parameters are required:
[Profibus-DP]
Setup/
Fieldbus parameter
Protocol
Profibus-DP
Profibus-DP Address
1 Enter address.
[DeviceNet]
Setup/
Fieldbus parameter
Protocol
DeviceNet
DeviceNet baudrate
500k Press
to select 500, 250 or 125 k.
DeviceNet MAC-ID
1 address 1 ... 62.
[ProfiNet I/O]
Setup/
Fieldbus parameter
Fieldbus protocol
ProfiNet I/O
Use DHCP
off Select
on/off.
IP address
192.168.1.1 Enter IP address.
Subnet mask
255.255.255.0 Enter Subnet mask.