11 parameter input, Parameter input, Pr 5220 instrument manual commissioning – Rice Lake PR5220 Ethernet Transmitter User Manual

Page 93

Advertising
background image

PR 5220 Instrument Manual

Commissioning

Sartorius

EN-93

4.6.11

Parameter Input

Weighingpoint/WP A/

Calibration

Ambient conditions

Very stable cond.

W

& M

none

Standstill time

0.50 s

Standstill range

1.00 d

Tare timeout

2.5 s

Zeroset range

50.00 d

[Ambient conditions]
This parameter is used to define the ambient conditions of the scale.
Possible values: very stable condition, Stable condition, Unstable condition, Very unstable condition

[W

& M]

Setting for operation in Legal-for-Trade mode.
Select [none], [OIML], [NTEP] (für USA) or [NSC] (for Australia), see Chapter 4.4.13.1.

[Standstill time]
The parameters [Standstill time] and [Standstill range] can be used to define the stability of the scale (stable
balance).
The [Standstill time] parameter is entered in seconds. The permitted range is 0.00…2 s.
The time can be entered from 0.00 to 2.00 seconds, but makes up at least one measurement time.

[Standstill range]
As long as the weight fluctuations remain within this range, the device is determined to be stable.
The [Standstill range] parameter is entered in “d.” The permitted range is 0.01…10.00 d.
For use in Legal-for-Trade mode, <1 d must be selected.

[Tare timeout]
Timeout for a tare/zeroset command that cannot be executed (e.g., due to mechanical instability of the scale,
incorrect filter setting, resolution too high, standstill condition too strict).
This parameter is given in seconds. The permitted range is 0.0…<2.5>…25 s.
At 0.0 s taring is only carried out when the scale is already stable.

[Zeroset range]
Determine a ±range around the zero point determined by the dead load during calibration; within this range
-

the displayed gross weight can be set to zero by pressing the zero-setting key (or by a corresponding

external command), and

-

automatic zero tracking is active.

Setting range: 0.00…10000.00 d
For use in Legal-for-Trade mode a value <2% of Max must be entered. Example: 60 d for 3000 e, class III.