Commissioning, Pr 5220 instrument manual – Rice Lake PR5220 Ethernet Transmitter User Manual
Page 94
Commissioning
PR 5220 Instrument Manual
EN-94
Sartorius
Weighingpoint/WP A/
Calibration
Zerotrack range
0.25 d
Zerotrack step
0.25 d
Zerotrack time
1.0 s
Overload
9 d
Min
20 d
Range mode
Single range
[Zerotrack range]
Range within which the automatic zerotrack is balanced.
Setting range: 0.25…10000.00 d
For use in Legal-for-Trade mode a value of 0.25 d must be entered.
[Zerotrack step]
If a weight change exceeds the adjusted value, automatic tracking does not function any more.
Setting range for automatic tracking increments: 0.25... 10 d
For use in Legal-for-Trade mode a value of 0.25 d must be entered.
[Zerotrack time]
Time interval for automatic zero tracking.
Setting range: 0.0…25 s
At 0.0 s the tracking is switched off. For use in Legal-for-Trade mode a value of 1 s must be entered.
[Overload]
Weighing range above the maximum capacity (Max) without error message.
Setting range: 0…999900 d. For use in Legal-for-Trade mode a value of max. 9 d = e must be set.
[Min]
Minimum weight at which a print command can be triggered.
Setting range: 0…999900 d. For use in Legal-for-Trade mode a value of at least 20 d must be set.
[Range mode]
For scale range selection, see Chapter 4.4.13.2.