Pr 5220 instrument manual commissioning – Rice Lake PR5220 Ethernet Transmitter User Manual

Page 89

Advertising
background image

PR 5220 Instrument Manual

Commissioning

Sartorius

EN-89

?

Corner correction will be reset.

Yes

No

Press [No] for no corner correction reset.
Press [Yes] to reset the corner correction and calibrating continues.

Weighingpoint/WP A/

Calibration

Parameter input:

Number of vessel feet

4 Number of vessel feet (in the case, 4).

Local gravity

9.81379 m/s

2

Local gravity

Max

10000 d

200.000 t Max. weight

Scale interval

1 d

0.020 t Scale interval

Dead load

Dead load: empty scale

CAL weight

Calibrating weight

Corner correction

Corner correction


[Number of vessel feet]
Enter number of vessel feet.

[Local gravity]
Enter local gravity (in this case: Hamburg 9.81379 m/s

2

); see e.g.

http://www.ptb.de/cartoweb3/SISproject.php.

[Max]
The load cell capacity is suggested as Max (E

max

* number of load cells).

The maximum capacity (Max) determines the maximum measured weight without dead load. Normally, the
selected Max must be smaller than the load cell capacity (nominal load * number of load cells) – dead load, in
order to prevent overloading the load cells.
Adapt the value for Max. When entering Max, the required number of digits behind the decimal point must be
taken into account. Press

to switch units.

[Scale interval]
Select the scale interval (1, 2, 5, 10, 20 or 50). The number of intervals is calculated and fed back
automatically.