Commissioning, Pr 5220 instrument manual – Rice Lake PR5220 Ethernet Transmitter User Manual
Page 58
Commissioning
PR 5220 Instrument Manual
EN-58
Sartorius
Possible Error messages
This message displays, if the maximum capacity is too high
(the calculated input voltage for the specified maximum capacity
exceeds 36 mV).
Subsequent changing of the maximum capacity is possible; if you
decrease the capacity, a message is displayed if the new maximum
capacity is lower than the test load ([Calibrated at]).
This message displays, if the selected resolution is to low, e.g. 5 kg.
The selected resolution is so high that less than 0.8 internal counts per
scale interval (d) or 0.5 µV/e for legal-for-trade acc. to OIML/NSC are
available.
This message displays, if the maximum capacity is not an integer
multiple of the scale interval.
After you press [OK], the input value for the maximum capacity is canceled.