Installing the Instrument

PR 5220 Instrument Manual

EN-26

Sartorius

3.1.4

Optocoupler Inputs

The 3 optocoupler inputs have one common potential (GND) for the input group that is separated from the
common potential of the output group.

Connection method 4-pin plug-in terminal block
Cable

Shielded, max. 50 m

Number of outputs

3

Input signal

External supply required
10…28 V DC for 'high' level
0…5 V DC for 'low' level

Input voltage

Max. 28 V DC

Input current

<11 mA @ 24 V DC
<5 mA @ 12 V DC

Electrical isolation

Yes; a common minus potential for the group
of 3 inputs

Example: contact input connection

When a voltage

≥10 V DC is applied to the terminals (in the

example:1-GND), input 1 is active (true).