Commissioning

PR 5220 Instrument Manual

EN-108

Sartorius

4.9.2

Configuring Digital Inputs

An action both for signal change from 0 to 1 (on) and from 1 to 0 (off) can be determined for each of the three
inputs. Digital inputs can be linked with conditions that must be met before an action can be started.

Press

and select [Digital I/O parameter] to open the configuration menu.

Setup/

Digital I/O parameter

Output 1

marker bit 1

X64=1

Output 2

marker bit 2

X65=1

Output 3

marker bit 3

X66=1

Input 1 on

set tare

X113=1

Condition

no condition

-----

Input 1 off

-no action-

Input 2 on

-no action-

Input 2 off

-no action-

Input 3 on

-no action-

Input 3 off

-no action-

BCD out

Gross

1. Determining an action
Determine the action for the rising edge of input 1 under [Input 1 on] from the following list (here: When the
input signal changes from 0 to 1, a tare command is generated).
Accordingly, an action for the falling edge can be determined.
Selection list for actions of the inputs [Input 1/2/3 on/off]

Function

SPM bit Description

-no action-

-----

No function

set marker 1

X64=1

Set marker 1

set marker 2

X65=1

Set marker 2

set marker 3

X66=1

Set marker 3

select net

X72=1

Select net

set zero

X112=1

Set zero

set tare

X113=1

Set tare

reset tare

X114=1

Reset tare

set test

X115=1

Activate the analog test

reset test

X116=1

Finish the analog test

reset PWF

X117=1

Reset power fail

set fixtare

X118=1

Set fixtare (use the value in address D31 as a tare value)

get fixtare

X119=1

Save gross value as fixtare in address D31

set print

X120=1

Activate a print order

clr marker 1

X64=0

Clear marker 1

clr marker 2

X65=0

Clear marker 2

clr marker 3

X66=0

Clear marker 3

select gross

X72=0

Save the gross weight in address D11