7 xbpi calibration with the user weight, Xbpi calibration with the user weight, Commissioning – Rice Lake PR5220 Ethernet Transmitter User Manual

Page 78: Pr 5220 instrument manual

Advertising
background image

Commissioning

PR 5220 Instrument Manual

EN-78

Sartorius

4.5.7

xBPI Calibration with the User Weight

Prerequisites:
-

The xBPI protocol has been selected (see Chapter 4.5.1).

-

The weighing point ‚xBPI-scale’ has been selected (see Chapter 4.5.4).

-

The platform configuration has been executed (see Chapter 4.5.3).

-

The setting in the following menu has been done: [Weighingpoint A-xBPI-Scale]-[Setup]:

[Configuration]-[Confirming adjust.]: [manual] (see Chapter 4.5.5).

-

The communication between instrument and platform is active.

Weighingpoint/

xBPI-Scale Setup

Calibration

Open with .

Configuration

Select group of specification

Show device info

Weighingpoint/

xBPI-Scale Setup

Dead load

Set

Delete

Span

Adjust with user weight

Select with

and enter with

.

Adjust with auto weight

Enter user weight

The previously stored user weight is displayed.

2000 g The weight can be changed.

Following window appears:

Weighingpoint/

xBPI-Scale Setup

Calibration status

Load to small Calibration progress without weight.

Cal-Target

-2000 g

--

0.01 g

The following window is displayed after applying the weight:

Weighingpoint/

xBPI-Scale Setup

Calibration status

Difference display

Cal-Delta

-0.3 g

--

1999.75 g The weight is displayed in high-resolution (10x).

Select [Accept].

Accept

ResError

Abort