Extended Functions
PR 5220 Instrument Manual
EN-152
Sartorius
6.
Click the relevant file in the ‘Explorer’ with the mouse and drag it into
window [Select
]
(or use ‘Copy and paste’).
7.
Select [Network].
8.
Fill in the device IP address.
9.
Click [Flash] to start the procedure.
The first three LEDs blink in a pattern on the device for 30 seconds.
10.
Press the reset key on the device a short time, within these 30 seconds
using a pen tip or paper clip with a diameter of 1.0 mm.
Note:
The flash procedure will not start if the reset key is not pressed within the 30-
second time period.
During the flash procedure, LEDs 1–4 blink in a pattern.
The device is restarted after the software has loaded.
The next file can be loaded as described in steps 6–10.
Note:
If the device cannot start because the software has not loaded correctly, the
first three LEDs blink in a pattern to prompt an update.