13 error messages, 1 measuring circuit error messages, Error messages – Rice Lake PR5220 Ethernet Transmitter User Manual

Page 154: Measuring circuit error messages, Pr 5220 instrument manual

Advertising
background image

Error Messages

PR 5220 Instrument Manual

EN-154

Sartorius

13

Error Messages

13.1

Measuring Circuit Error Messages

The weighing electronics can generate error messages, which are output on the weight display.

VNC Text

Error

Possible Cause

Arith err

Internal arithmetic overflow

-

Faulty calibration values

Overload

Input voltage is higher than
Max + (x d)

-

To much weight on the scale

Ext. meas.device error Internal weighing point:

Input signal is higher than the
permissible range of 36 mV.

-

Defective load cell

-

Cable break

xBPI scale:
Cannot read weight values from scale.

-

Cable break

-

Internal scale error

No sense voltage

No sense voltage

-

Load cells not connected

-

Sense line or supply line is interrupted.

-

Wrong polarity or sense voltage is low.

Negative input

Negative input

-

Wrong polarity of load cell signal.

-

Wrong polarity of supply voltage.