13 error messages, 1 measuring circuit error messages, Error messages – Rice Lake PR5220 Ethernet Transmitter User Manual
Page 154: Measuring circuit error messages, Pr 5220 instrument manual
Error Messages
PR 5220 Instrument Manual
EN-154
Sartorius
13
Error Messages
13.1
Measuring Circuit Error Messages
The weighing electronics can generate error messages, which are output on the weight display.
VNC Text
Error
Possible Cause
Arith err
Internal arithmetic overflow
-
Faulty calibration values
Overload
Input voltage is higher than
Max + (x d)
-
To much weight on the scale
Ext. meas.device error Internal weighing point:
Input signal is higher than the
permissible range of 36 mV.
-
Defective load cell
-
Cable break
xBPI scale:
Cannot read weight values from scale.
-
Cable break
-
Internal scale error
No sense voltage
No sense voltage
-
Load cells not connected
-
Sense line or supply line is interrupted.
-
Wrong polarity or sense voltage is low.
Negative input
Negative input
-
Wrong polarity of load cell signal.
-
Wrong polarity of supply voltage.