background image

PR 5220 Instrument Manual

Commissioning

Sartorius

EN-55

4.4

Calibration Weighing Point ‚Internal A’

Calibration using weights, mV/V or load cell data can be done using the VNC program. During calibration, the
instrument must be set to gross weight display (reset tare, if necessary).

For a ’legal-for-trade’ application, set the mode under

-[Weighingpoint]-[Calib]- [Param] to [W

&M] before

starting calibration; see Chapter 4.4.13.1. Select [New] to go to the maximum capacity [Max]
(see chapter 4.4.3), select the scale interval and determine the dead load. Now calibrate the maximum capacity
by applying a calibration weight, in mV/V or with load cell data. After determining the test value, the menu
can be closed as described in Chapter 4.4.12, in order to save the new settings. Calibration data can be
protected by the CAL switch (see Chapter 4.1.1), which must be sealed in the closed (write-protected) position
for ‘legal-for-trade’ applications.

4.4.1

Displaying Calibration Data

When the CAL switch is closed, the following message is displayed; only
data display possible with [Param]:


With [View], the calibration data can be displayed, but not changed.

Weighingpoint/WP A/

View Calibration

Max 3000 d 3000 kg

Number of scale intervals and max. capacity

Scale interval 3000 d 1 kg

Scale interval

Deadload at 0.00 kg 0.000000

Dead load in weight and mV/V

Max at 3000.00 kg 1.000000

Weight and mV/V for maximum capacity

Calibrated at 3000.00 kg 1.000000

Test load* and corresponding mV/V

Sensitivity 833.33

4.000000

Number of internal counts and voltage

per scale interval

Param

The calibration data and parameters (press [Param]) are displayed in the format entered/determined during
calibration.
* After input with mV/V, the maximum capacity and the mV/V value entered are displayed.

4.4.1.1

Increased Resolution (10-Fold)

In the

-[Weighingpoint]-[Calib] menu, the weight is displayed with 10-fold resolution (also with the CAL

switch closed) when you press the key

, and marked as an invalid weight with

above the weight unit.

After 5 s, the display returns to normal resolution.

You can press the key

to return to normal display immediately.