PR 5220 Instrument Manual
Commissioning
Sartorius
EN-105
Condition
SPM bit
Description
actual diginp1
X00=1
Digital input 1: active
actual diginp2
X01=1
Digital input 2: active
actual diginp3
X02=1
Digital input 3: active
actual limit 1
X16=1
Limit signal 1: active
actual limit 2
X17=1
Limit signal 2: active
actual limit 3
X18=1
Limit signal 3: active
ADC error
X32=1
General error in the weighing point
above Max
X33=1
Weight above Max
overload
X34=1
Weight above Max plus the ‘overload’ value
below zero
X35=1
Weight below zero
center zero
X36=1
Weight within ¼ d of zero
inside ZSR
X37=1
Weight within zero-setting range
standstill
X38=1
Mechanical stability of the scale
out
X39=1
Weight below zero or above Max
command error
X48=1
For internal use only.
command busy
X49=1
For internal use only.
power fail
X50=1
Set after power-on (=power failure)
test active
X56=1
Analog test was started.
cal active
X57=1
For internal use only.
tare active
X58=1
Instrument is tared.
marker bit 1
X64=1
Marker bit 1 set, after power-on the markers are set to ‘0’.
marker bit 2
X65=1
Marker bit 2 set, after power-on the markers are set to ‘0’.
marker bit 3
X66=1
Marker bit 3 set, after power-on the markers are set to ‘0’.