PR 5220 Instrument Manual

Specifications

Sartorius

EN-161

14.4

Weighing Electronics

The weighing electronics are suitable for connection of strain-gauge load cells.

14.4.1

Load Cells

Load cell type

Strain gauge, 6 or 4-wire connection possible

Supply voltage

U= 12V DC for I

max

= 160 mA

for up to 8 load cells of 650

Ω each

or 4 load cells of 350

Ω each

Sense voltage monitoring

Sense voltage below +4VDC -- -4VDC is detected; can be switched off

Max. load

≥75 Ω

Cable length

Max. 500 m

14.4.2

Principle

Principle

DC voltage, delta-sigma converter, ratiometric to the load cell supply voltage

Conversion/measurement time

5, 10, 20, 40, 80, 160, 320, 640, 960, 1200, 1600 ms

Digital filter

Selectable, active 4th order (low-pass),
Characteristic: Bessel, aperiodic, Butterworth, Tschebyscheff

Configurable cut-off frequency

14.4.3

Accuracy and Stability

Accuracy class

≤10000 e (Cl. III) acc. to OIML R76 / EN 45501

Min. measuring signal (OIML)

0.5µV/e, i.e. 3 mV for 6000 e, 5 mV for 10000 e

Linearity error:

<0.002 %

Zero stability error (TK

0

)

<0.02 µV/K RTI <= 0.004%/10K at 1mV/V

Span stability error (TK

Spn

)

< ±2 ppm/K

14.4.4

Sensitivity

Sensitivity

0.5 µV/e @ 10000e (Cl. III) OIML R76,

Resolution

7.5 million counts at 3mV/V, not legal for trade

Input voltage (input signal + dead load)

0 ... max. 36 mV DC, symmetrical to 0

Dead load range

36 mV DC (max. input signal); input/ calibration via software