Fieldbus interface, Pr 5220 instrument manual – Rice Lake PR5220 Ethernet Transmitter User Manual

Page 144

Advertising
background image

Fieldbus Interface

PR 5220 Instrument Manual

EN-144

Sartorius

Register 30, 31: Fixed Values (Read/Write)

Register 30

Fixed value for analog output; value (num) 0... 20000 corresponds to 20 mA

Register 31

Fixed value for fixtare; see also SetFixTare, GetFixTare ( see page 142)

Register 80 ... 89: State-Controlled Action Bits (Write)
For setting bits, see page 138.
Only setting and resetting of single bits is possible.
When changing a bit from 0 to 1, the corresponding action starts. After handling the command, the bit must
be reset. Application: the master writes cyclically.
The bit is set as Write_Value_Select with the specified number (see page 138); the bit is reset at the specified
number +128.

Register 80 SetZero

Set the gross weight to zero

Register 81 SetTare

Execute taring

Register 82 ResetTare

Reset tare

Register 83 SetTest

Start the ADC test

Register 84 ResetTest

Finish the ADC test

Register 85 ResetPwf

Reset the PowerFail bit (Register 1; the bit was set after power on)

Register 86 SetFixTare

Taring with weight in numerical address D31 'Fixtare'

Register 87 GetFixTare

The current gross weight is copied into numerical address D31.

Register 89 ResetError

The CmdError error bit is reset

Register 112 ... 121: Transition-Controlled Action Bits (Write)
For setting bits, see page 138.
As soon as the bit was set, it is reset internally and the operation is handled; this is transition-controlled (for
writing once).
The bit is set as Write_Value_Select with the specified number (see page 138).

Register 112

SetZero

Register 113

SetTare

Register 114

ResetTare

Register 115

SetTest

Register 116

ResetTest

Register 117

ResetPwf

Register 118

SetFixTare

Register 119

GetFixTare

Register 121

ResetError