PR 5220 Instrument Manual

Commissioning

Sartorius

EN-57

4.4.2.2

Changing the Calibration

Note:

[Modify] may be used only for minor changes (e.g. changing the dead load, adapting mV/V
values for dead load and/or Max); otherwise, always use [New].

Open the menu via

-[Weighingpoint]-[Calib]

-[Modify].

Weighingpoint/WP A/

Calibration

Max

3000 d

3000 kg

Scale interval

3000 d

1 kg

Deadload at

1.07 kg

0.000358

Max at

3000.00 kg

1.000000

Calibrated at

3000.00 kg

1.000000

Sensitivity

833.33

4.000000

by load

by mV/V

Test

For setting a new value for Dead load, press

to select [Deadload] and either enter a new

value

with [by mV/V] or discharge the scale/hopper and press [by load].

When closing the menu with

you are prompted whether the menu should

be closed without calculation of the test value:
Reply [Yes] to close the menu.

4.4.3

Determining the Maximum Capacity (Max)

The maximum capacity (Max) determines the maximum weight without dead load of the weight to be
measured and the displayed number of digits behind the decimal point. Normally, Max is less than the load cell
capacity (nominal capacity x number of load cells).
Permissible values are:

[Max] from 0.00010 to 999999, with

in kg, t, g or lb

Max must be an integer multiple of the scale interval. It may have up to 6 digits and is entered as a numeric
value with or without decimal point.
Note:

If the liniarization is active (see Chapter 4.4.10):
After selection of the line ‚Max at’ the following tip is displayed:

Can not be changed here

while linearization is active

Weighingpoint/WP A/

Calibration

Max at

3000 d

3000 kg

The weight unit can be changed from kg into t, g or lb by pressing

.

After pressing

or

confirmation of the change is displayed with:

Setting Max