Commissioning
PR 5220 Instrument Manual
EN-92
Sartorius
4.6.10
Finishing/Saving the Calibration
Press
to complete the calibration.
Unless all data were determined during calibration using [New] (e.g. dead load
not set / entered), the following prompt is shown:
After replying [Yes] to confirm and pressing
, another prompt is displayed:
Press [Save] to save changes in calibration data. If you press [Undo], changes
are not saved and the display returns to the selection menu for the weighing
points.
The Confirmation is displayed with:
Saving calibration
After quitting the menu, the following message is displayed:
Exit calibration
After finishing the calibration, set the CAL switch to the closed position; see also Chapter 4.1.1.