8 xbpi calibration with automatic weight detection, Xbpi calibration with automatic weight detection, Pr 5220 instrument manual commissioning – Rice Lake PR5220 Ethernet Transmitter User Manual
Page 79
PR 5220 Instrument Manual
Commissioning
Sartorius
EN-79
The data are saved and the instrument generates a corresponding message:
Weighingpoint/
xBPI-Scale Setup
Calibration status
complete
Net=Grs
2000 g
--
2000.01 g The weight is displayed in high-resolution (10x).
Leave with
.
Accept
ResError
Abort
4.5.8
xBPI Calibration with Automatic Weight Detection
Prerequisites:
-
The xBPI protocol has been selected (see Chapter 4.5.1).
-
The weighing point ‚xBPI-scale’ has been selected (see Chapter 4.5.4).
-
The platform configuration has been executed (see Chapter 4.5.3).
-
The setting in the following menu has been done: [Weighingpoint A-xBPI-Scale]-[Setup]:
[Configuration]-[Confirming adjust.]: [manual] (see Chapter 4.5.5).
-
The communication between instrument and platform is active.
Weighingpoint/
xBPI-Scale Setup
Calibration
Open with .
Configuration
Select group of specification
Show device info
Weighingpoint/
xBPI-Scale Setup
Dead load
Set
Delete
Span
Adjust with user weight
Adjust with auto weight
Select with
and enter with
.
Following window appears:
Weighingpoint/
xBPI-Scale Setup
Calibration status
Load to small Calibration progress without weight.
Cal-Target
-10000.0 g The weight is displayed in high-resolution (10x).
--
-0.02 g