Super Systems 9205 Series User Manual

Page 110

Advertising
background image

Series 9205 Operations Manual Rev A

109

Decimal Point Location

This is the decimal point location value. This will affect the PV value and the location of the decimal when it

is displayed. Clicking on this value will display an input box from which the user can select a new value.

The range is 0 to 4.

Open Input

This is the open TC value. Clicking on this value will toggle between up scale, down scale, one trip point,

and two trip points.

Input Offset

The input offset value is algebraically added to the input value to adjust the input curve on read-out. The

range is –5000 to 5000.


TRIP POINT EXPLANATION

Setting a trip point will force the value that the controller uses for calculations to a certain value as

assigned by the operator. Once the Trip Point Setpoint is reached, the controller will begin reading the

value as the Trip Point Force Value, regardless of what the actual value is inside the furnace. The Trip Point

Direction allows the operator to choose whether the controller will alter its reading when the trip point is

either above or below the setpoint.

Trip Point 1 Setpoint

This is the trip point 1 setpoint value in °F. The trip point is used as a way to ensure that the probe is still

working properly and that there is not a problem with the furnace that could ruin the load. If the probe

takes a measurement in conflict with the trip point, the reading will automatically be forced to the Trip

Point Force Value. This will cause the system to set off an alarm so that the operator will be aware there is

a problem. The range is –32768 to 32768.

Trip Point 1 Force Value

This is the trip point 1 force value. This value is used when the probe begins reading values in conflict with

the trip point. Although the probe will still be taking measurements, the 9205 controller will use the force

value as the input so that an alarm will be set off. The range is –32768 to 32768.

Trip Point 1 Direction

This is the trip point 1 direction. The options are: input above setpoint or input below setpoint. If input

above setpoint is chosen, then the force value will be used when the probe begins measuring above the trip

point setpoint. If input below setpoint is chosen, then the force value will be used when the probe begins

measuring below the trip point setpoint.

Trip Point 2 Setpoint

This is the trip point 2 setpoint value. The range is –32768 to 32768.

Trip Point 2 Force Value

This is the trip point 2 force value. The range is –32768 to 32768.

Trip Point 2 Direction

This is the trip point 2 direction. The options are: input above setpoint or input below setpoint.

High Input Limit Setpoint

This is the setpoint for the high input limit. The setpoint cannot be assigned any value above this. The

range is –32768 to 32768.

Advertising
This manual is related to the following products: