Configuration – Super Systems 9205 Series User Manual

Page 62

Advertising
background image

Series 9205 Operations Manual Rev A

61

would be set to 3.6. If 1700 °F is being sourced, and the “Value” is showing 1702.3 °F, then the “Offset”

would be set to -2.3.

Click on the Calibrate button to begin the calibration.

Wait 120 seconds and verify with a source calibration device with the correct T/C type.

Note: During a

normal calibration procedure, the user should zero and span all of the inputs first and then perform a cold

junction calibration, if necessary

.

It is recommended after each Cold Junction calibration to source a value in and check what the displayed

temperature reading is.

Click on the Return button to close down the screen.

Configuration

This option will allow the user to set some general configuration settings.

The list of configurable items is:

Log Maintenance

Programmer Alarm Text

Event Text

Compact Database

Screen

Datalogging Setup

Repair Database

Device Configuration

Recipe Configuration

Product Activation

Alarm Text

Time Synchronization

Probe Management

Maintenance

Read/Write Data

Log Maintenance

This option will allow the user to purge, or delete, log files,

starting from a specific date. All log files from on or before

the selected date will be purged. The user can purge log

files from the Alarm Log, Event Log, or System Log.

NOTE: The touch screen may require a login to access this
menu.





Advertising
This manual is related to the following products: