SSi Analog Input Setup

The SSi Analog Input Setup menu option allows the user an

input selection of three inputs per module, and three input

corrections per module. There are eight modules

available. It is configurable for voltage of T/C (universal

input), and it is typically used for Load T/Cs and Auxiliary

Flow Meters. Before connecting your input source to the

terminals, please verify that the input type is set up

correctly. If the Input Type is not correct, do not connect

the input source to the terminals, as damage can occur.

Please consult SSi by calling (513) 772-0060 before making

any changes.

Input Type 0 – Input Type 2

This will select the input type for the module. The options are:

B

NNM

160 mV

C

R

80 mV

E

S

40 mV

J

T

20 mV

25.6 volts**

K

2.56 volts

4-20 mA/124

12.8 volts**

N

1.28 volts

4-20 mA/62

** - When the specified input type is selected, a jumper located inside the case will need

to be placed on that specific input for reading this selection. If jumper is not placed on

input, then damage could occur to the board.

Input 0 Correction – Input 2 Correction

This option will set a correction curve for the input. The options for the input corrections are: not used,

Curve 1 – Curve 3.

Input 1 Open T/C – Input 3 Open T/C

This option will allow the user to set the direction of the open T/C for each input. The options are: Up Scale

or Down Scale.