Source]:sweep:step:stop:frequency:step – RIGOL DSG3000 Series User Manual

Page 115

Advertising
background image

Chapter 2 Command System

RIGOL

DSG3000 Programming Guide

2-95

[:SOURce]:SWEep:STEP:STOP:FREQuency:STEP

Syntax

[:SOURce]:SWEep:STEP:STOP:FREQuency:STEP <value>
[:SOURce]:SWEep:STEP:STOP:FREQuency:STEP?

Description

Set the stop frequency step of the sweep.
Query the stop frequency step of the sweep.

Parameter

Name

Type

Range

Default

<value>

Real

10mHz to 1GHz

100MHz

Explanation

 When <value> is set in "Number" form, the default unit is Hz. Besides,

<value> can also be set in "Number + Unit" form, for example, 3kHz.

 The default unit of the return value is Hz.

 After setting the stop frequency step, you can use the up/down direction keys

or knob to modify the stop frequency at the current step. At this point, you can

query or set the stop frequency using the

[:SOURce]:SWEep:STEP:STOP:FREQuency

command.

Return Format

The query returns the stop frequency step of the sweep, for example, 3000.

Example

:SWE:STEP:STOP:FREQ:STEP 3kHz
:SWE:STEP:STOP:FREQ:STEP?

Related

Command

[:SOURce]:SWEep:STEP:STOP:FREQuency