Source]:pm:deviation, Source]:pm:deviation:step[:increment – RIGOL DSG3000 Series User Manual

Page 81

Advertising
background image

Chapter 2 Command System

RIGOL

DSG3000 Programming Guide

2-61

[:SOURce]:PM:DEViation

Syntax

[:SOURce]:PM:DEViation <value>
[:SOURce]:PM:DEViation?

Description

Set the phase deviation of ØM.
Query the phase deviation of ØM.

Parameter

Name

Type

Range

Default

<value>

Real

0rad to 20rad

6.28rad

Explanation

 When <value> is set in "Number" form, the default unit is rad. Besides,

<value> can also be set in "Number + Unit" form, for example, 2rad.

The default unit of the return value is rad.

 After setting the phase deviation, you can use the up/down direction keys or

knob to modify the phase deviation at the current step. At this point, you can

query or set the current step using the

[:SOURce]:PM:DEViation:STEP[:INCRement]

command.

Return Format

The query returns the phase deviation of ØM, for example, 2.000000.

Example

:PM:DEV 2
:PM:DEV?

Related

Command

[:SOURce]:PM:DEViation:STEP[:INCRement]

[:SOURce]:PM:DEViation:STEP[:INCRement]

Syntax

[:SOURce]:PM:DEViation:STEP[:INCRement] <value>
[:SOURce]:PM:DEViation:STEP[:INCRement]?

Description

Set the ØM phase deviation step.
Query the ØM phase deviation step.

Parameter

Name

Type

Range

Default

<value>

Real

0.01rad to 10rad

1rad

Explanation

 When <value> is set in "Number" form, the default unit is rad. Besides,

<value> can also be set in "Number + Unit" form, for example, 3rad.

 The default unit of the return value is rad.

 After setting the phase deviation step, you can use the up/down direction

keys or knob to modify the phase deviation at the current step. At this point,

you can query or set the phase deviation using the

[:SOURce]:PM:DEViation

command.

Return Format

The query returns the phase deviation step (3.000000).

Example

:PM:DEV:STEP 3
:PM:DEV:STEP?

Related

Command

[:SOURce]:PM:DEViation