Source]:level:limit – RIGOL DSG3000 Series User Manual

Page 63

Advertising
background image

Chapter 2 Command System

RIGOL

DSG3000 Programming Guide

2-43

[:SOURce]:LEVel:LIMit

Syntax

[:SOURce]:LEVel:LIMit <value>
[:SOURce]:LEVel:LIMit?

Description

Set the limit of the RF output amplitude.
Query the limit of the RF output amplitude.

Parameter

Name

Type

Range

Default

<value>

Real

-140dBm to 25dBm

25dBm

Explanation

 When <value> is set in "Number" form (for example, 2), the default unit is

dBm. When it is set in "Number + Unit" form (for example, 2dBm), the limit

displayed in the RF signal generator interface is related to the setting of Level

Unit.

When the level unit is "dBm", it is displayed as 2.00dBm;

When the level unit is "dBmV", it is displayed as 48.99dBmV;

When the level unit is "dBuV", it is displayed as 108.99dBuV;

When the level unit is "Volts", it is displayed as 281.50mV;

When the level unit is "Watts", it is displayed as 1.58mW.

 The default unit of all the return values is dBm.

 After setting the amplitude limit, you can use the up/down direction keys or

knob to modify the limit at the current step. At this point, you can query or set

the current step using the

[:SOURce]:LEVel:LIMit:STEP

command.

Return Format

The query returns the limit of the RF output amplitude, for example, 5.60.

Example

:LEV:LIM 5.6
:LEV:LIM?

Related

Command

[:SOURce]:LEVel:LIMit:STEP

Advertising