Source]:pulm:dwidth:step, Source]:pulm:mode – RIGOL DSG3000 Series User Manual

Page 89

Advertising
background image

Chapter 2 Command System

RIGOL

DSG3000 Programming Guide

2-69

[:SOURce]:PULM:DWIDth:STEP

Syntax

[:SOURce]:PULM:DWIDth:STEP <value>
[:SOURce]:PULM:DWIDth:STEP?

Description

Set the #2 pulse width step of the double pulse modulating signal.
Query the #2 pulse width step of the double pulse modulating signal.

Parameter

Name

Type

Range

Default

<value>

Real

10ns to 10s

100us

Explanation

 When <value> is set in "Number" form, the default unit is s. Besides, <value>

can also be set in "Number + Unit" form, for example, 3000ms.

 The default unit of the return value is s.

 After setting the #2 pulse width step, you can use the up/down direction keys

or knob to modify the width at the current step. At this point, you can query or

set the #2 pulse width using the

[:SOURce]:PULM:DWIDth

command.

Return Format

The query returns the #2 pulse width step of the double pulse modulating signal, for

example, 3.

Example

:PULM:DWID:STEP 3000ms
:PULM:DWID:STEP?

Related

Command

[:SOURce]:PULM:DWIDth

[:SOURce]:PULM:MODE

Syntax

[:SOURce]:PULM:MODE DOUBle|SINGle|TRAin
[:SOURce]:PULM:MODE?

Description

Set the mode of pulse modulation.
Query the mode of pulse modulation.

Parameter

Name

Type

Range

Default

DOUBle|SINGle|TRAin Discrete

DOUBle|SINGle|TRAin

SINGle

Explanation

 DOUBle: set the pulse mode to "Double" to enable the double pulse

modulation mode.

 SINGle: set the pulse mode to "Single" to enable the single pulse modulation

mode.

 TRAIn: set the pulse mode to "Train" to enable the train modulation mode. At

this point, you can use the

[:SOURce]:PULM:TRAin:LIST:INSErt

command to

define the desired pulse train.

 This command is invalid when "Ext" modulation source is selected.

Return Format

The query returns SINGLE, DOUBLE or TRAIN.

Example

:PULM:MODE DOUB
:PULM:MODE?

Related

Commands

[:SOURce]:PULM:SOURce
[:SOURce]:PULM:TRAin:LIST:INSErt