[:SOURce]:LFOutput:SWEPtsine:TRIGger

Syntax

[:SOURce]:LFOutput:SWEPtsine:TRIGger AUTO|KEY|BUS|EXT
[:SOURce]:LFOutput:SWEPtsine:TRIGger?

Description

Set the trigger mode of sweep-sine.
Query the trigger mode of sweep-sine.

Parameter

Name

Type

Range

Default

AUTO|KEY|BUS|EXT

Discrete

AUTO|KEY|BUS|EXT

AUTO

Explanation

 AUTO: select "Auto" trigger mode. The RF signal generator fulfills the trigger

condition at any time and it will sweep according to the current setting once

"Swp-Sine" is selected.

 KEY: select "Key" trigger mode. If the sweep mode is set to "Cont", the

instrument performs a sweep each time Trigger at the front panel is pressed.
If the sweep mode is set to "Single", the instrument will perform a sweep and
then stop each time Trigger at the front panel is pressed after the

[:SOURce]:LFOutput:SWEPtsine:SINGle

command is sent and when the single

sweep condition is met.

 BUS: select "Bus" trigger mode. If the sweep mode is set to "Cont", the

instrument performs a sweep each time the

*TRG

or

:TRIGger:LFOutput[:IMMediate]

command is send. If the sweep mode is set

to "Single", the instrument performs a sweep and then stops each time the

*TRG

command is sent after the

[:SOURce]:LFOutput:SWEPtsine:SINGle

command is sent and when the single sweep condition is met.

 EXT: select "Ext" trigger mode. The RF signal generator accepts the trigger

signal input from the [TRIGGER IN] connector at the rear panel. If the sweep

mode is set to "Cont", the instrument performs a sweep each time a TTL pulse

signal with the specified polarity is received. If the sweep mode is set to

"Single", the instrument performs a sweep and then stops each time a TTL

pulse signal with the specified polarity is received after the

[:SOURce]:LFOutput:SWEPtsine:SINGle

command is sent and when the single

sweep condition is met.

Return Format

The query returns the trigger mode of sweep-sine, for example, KEY.

Example

:LFO:SWEP:TRIG KEY
:LFO:SWEP:TRIG?

Related

Commands

[:SOURce]:LFOutput:SHAPe
[:SOURce]:LFOutput:SWEPtsine:SINGle
[:SOURce]:SWEep:MODE
*TRG
:TRIGger:LFOutput[:IMMediate]