Source]:level:offset:step, Source]:level:sh:active – RIGOL DSG3000 Series User Manual

Page 66

Advertising
background image

RIGOL

Chapter 2

Command System

2-46

DSG3000 Programming Guide

[:SOURce]:LEVel:OFFSet:STEP

Syntax

[:SOURce]:LEVel:OFFSet:STEP <value>
[:SOURce]:LEVel:OFFSet:STEP?

Description

Set the step of the RF amplitude offset.
Query the step of the RF amplitude offset.

Parameter

Name

Type

Range

Default

<value>

Real

0.01dB to 200dB

10dB

Explanation

 When <value> is set in "Number" form, the default unit is dB. Besides,

<value> can also be set in "Number + Unit" form, for example, 20dB.

 The default unit of the return value is dB.

 After setting the step of the amplitude offset, you can use the up/down

direction keys or knob to modify the amplitude offset at the current step. At

this point, you can query or set the amplitude offset using the

[:SOURce]:LEVel:OFFSet

command.

Return Format

The query returns the step of the RF amplitude offset, for example, 20.00.

Example

:LEV:OFFS:STEP 20
:LEV:OFFS:STEP?

Related

Command

[:SOURce]:LEVel:OFFSet

[:SOURce]:LEVel:SH:ACTIve

Syntax

[:SOURce]:LEVel:SH:ACTIve

Description

Execute a sample hold operation immediately.

Explanation

This command is valid when the ALC function is disabled.

Related

Command

[:SOURce]:LEVel:ALC:MODE

Advertising