Source]:lfoutput:sweptsine:level:step, Source]:lfoutput:sweptsine:mode – RIGOL DSG3000 Series User Manual

Page 75

Advertising
background image

Chapter 2 Command System

RIGOL

DSG3000 Programming Guide

2-55

[:SOURce]:LFOutput:SWEPtsine:LEVel:STEP

Syntax

[:SOURce]:LFOutput:SWEPtsine:LEVel:STEP <value>
[:SOURce]:LFOutput:SWEPtsine:LEVel:STEP?

Description

Set the amplitude step of sweep-sine.
Query the amplitude step of sweep-sine.

Parameter

Name

Type

Range

Default

<value>

Real

1mV to 1V

0.5V

Explanation

 When <value> is set in "Number" form, the default unit is V. Besides, <value>

can also be set in "Number + Unit" form, for example, 1000mV.

 The default unit of the return value is V.

 After setting the amplitude step of sweep-sine, you can use the up/down

direction keys or knob to modify the amplitude at the current step. At this point,

you can query or set the amplitude of sweep-sine using the

[:SOURce]:LFOutput:SWEPtsine:LEVel

command.

Return Format

The query returns the amplitude step of sweep-sine, for example, 1.00.

Example

:LFO:SWEP:LEV:STEP 1000mV
:LFO:SWEP:LEV:STEP?

Related

Command

[:SOURce]:LFOutput:SWEPtsine:LEVel

[:SOURce]:LFOutput:SWEPtsine:MODE

Syntax

[:SOURce]:LFOutput:SWEPtsine:MODE CONTinue|SINGle
[:SOURce]:LFOutput:SWEPtsine:MODE?

Description

Set the mode of sweep-sine.
Query the mode of sweep-sine.

Parameter

Name

Type

Range

Default

CONTinue|SINGle

Discrete

CONTinue|SINGle

CONTinue

Explanation

 CONTinue: select "Cont" mode. When the trigger condition is met, the

instrument sweeps continuously at the current setting.

 SINGle: select "Single" mode. Send the

[:SOURce]:LFOutput:SWEPtsine:SINGle

command and the instrument

performs a sweep according to the current setting and then stops when the

trigger condition is met.

Return Format

The query returns CONT or SING.

Example

:LFO:SWEP:MODE CONT
:LFO:SWEP:MODE?

Related

Command

[:SOURce]:LFOutput:SWEPtsine:SINGle