Source]:sweep:sweep:trigger:type – RIGOL DSG3000 Series User Manual

Page 118

Advertising
background image

RIGOL

Chapter 2

Command System

2-98

DSG3000 Programming Guide

[:SOURce]:SWEep:SWEep:TRIGger:TYPE

Syntax

[:SOURce]:SWEep:SWEep:TRIGger:TYPE AUTO|KEY|BUS|EXT
[:SOURce]:SWEep:SWEep:TRIGger:TYPE?

Description

Set the trigger mode of the sweep period.
Query the trigger mode of the sweep period.

Parameter

Name

Type

Range

Default

AUTO|KEY|BUS|EXT

Discrete

AUTO|KEY|BUS|EXT

AUTO

Explanation

 AUTO: select "Auto" trigger mode. If the sweep mode is set to "Cont", the

instrument will start sweeping when a sweep type is selected. If the sweep

mode is set to "Single", the instrument will start a sweep and then stop after

the

[:SOURce]:SWEep:EXECute

command is sent and when the single sweep

condition is met.

 KEY: select "Key" trigger mode. If the sweep mode is set to "Cont", the

instrument starts a sweep each time Trigger at the front panel is pressed. If
the sweep mode is set to "Single", the instrument starts a sweep and then stops
each time Trigger at the front panel is pressed after the

[:SOURce]:SWEep:EXECute

command is sent and when the single sweep

condition is met.

 BUS: select "Bus" trigger mode. If the sweep mode is set to "Cont", the

instrument starts a sweep each time the

*TRG

or

:TRIGger[:SWEep][:IMMediate]

command is sent. If the sweep mode is set

to "Single", the instrument starts a sweep and then stops each time the

*TRG

or

:TRIGger[:SWEep][:IMMediate]

command is sent after the

[:SOURce]:SWEep:EXECute

command is sent and when the single sweep

condition is met.

 EXT: select "Ext" trigger mode. The RF signal generator receives the trigger

signal input from the [TRIGGER IN] connector at the rear panel. If the sweep

mode is "Cont", the instrument starts a sweep each time a TTL pulse signal with

the specified polarity is received. If the sweep mode is set to "Single", the

instrument starts a sweep and then stops each time a TTL pulse signal with the

specified polarity is received after the

[:SOURce]:SWEep:EXECute

command is

sent and when the single sweep condition is met.

Note: The above explanations are only valid when the trigger mode of each

sweep point within the sweep period is met.

 When executing the sweep operation, the priority of the required conditions is:

single sweep  trigger mode  point trigger mode.

Return Format

The query returns the trigger mode of the sweep, for example, AUTO.

Example

:SWE:SWE:TRIG:TYPE AUTO
:SWE:SWE:TRIG:TYPE?

Related

Commands

[:SOURce]:SWEep:EXECute
[:SOURce]:SWEep:MODE
[:SOURce]:SWEep:POINt:TRIGger:TYPE
*TRG
:TRIGger[:SWEep][:IMMediate]