Source]:pulm:train:list:insert, Source]:pulm:train:list:run – RIGOL DSG3000 Series User Manual

Page 95

Advertising
background image

Chapter 2 Command System

RIGOL

DSG3000 Programming Guide

2-75

[:SOURce]:PULM:TRAin:LIST:INSErt

Syntax

[:SOURce]:PULM:TRAin:LIST:INSErt <OnTime>,<OffTime>,<Repeat>

Description

Insert a row of pulse list data.

Parameter

Name

Type

Range

Default

<OnTime>

Real

20ns to 170s-20ns

500us

<OffTime>

Real

20ns to 170s-20ns

500us

<Repeat>

Integer

1 to 256

1

Explanation

 <OnTime>: set the duration of the positive pulse.

 <OffTime>: set the duration of the negative pulse.

 <Repeat>: set the repetition times of this pulse.

Example

:PULM:TRA:LIST:INSE 5.5s,2.5s,2 /*Insert a row of pulse value before the row

currently selected: positive delay 5.5s,

negative delay 2.5s and repetition times

2*/

[:SOURce]:PULM:TRAin:LIST:RUN

Syntax

[:SOURce]:PULM:TRAin:LIST:RUN

Description

Execute the pulse list currently edited.

Explanation

You can update the internal modulating signal as the current pulse list data using

this command.