Source]:iq:trigger:inhibit, Source]:iq:trigger:inhibit:step – RIGOL DSG3000 Series User Manual

Page 57

Advertising
background image

Chapter 2 Command System

RIGOL

DSG3000 Programming Guide

2-37

[:SOURce]:IQ:TRIGger:INHibit

Syntax

[:SOURce]:IQ:TRIGger:INHibit <value>
[:SOURce]:IQ:TRIGger:INHibit?

Description

Set the external inhibit of IQ wave table.
Query the external inhibit of IQ wave table.

Parameter

Name

Type

Range

Default

<value>

Integer

0 to 65535

0

Explanation

 After a trigger event occurs, the trigger signal is no longer received during the

external inhibit time.

 This command is valid only when selecting "Ext" trigger mode. <value>

describes the number of points (

i

N

). The actual time (

i

T

) can be obtained

according to the current sample rate (

a

S

):

a

i

i

S

N

T

=

.

 After the external inhibit is set, you can use the up/down direction keys or knob

to modify the inhibit at the current step. You can query or set the current step

using the

[:SOURce]:IQ:TRIGger:INHibit:STEP

command.

Return Format

The query returns the external inhibit, for example, 5000.

Example

:IQ:TRIG:INH 5000
:IQ:TRIG:INH?

Related

Commands

[:SOURce]:IQ:TRIGger:INHibit:STEP
[:SOURce]:IQ:TRIGger:MODe

[:SOURce]:IQ:TRIGger:INHibit:STEP

Syntax

[:SOURce]:IQ:TRIGger:INHibit:STEP <value>
[:SOURce]:IQ:TRIGger:INHibit:STEP?

Description

Set the external inhibit step.
Query the external inhibit step.

Parameter

Name

Type

Range

Default

<value>

Integer

1 to 10000

1

Explanation

After setting the external inhibit step, you can use the up/down direction keys or

knob to modify the external inhibit at the current step. At this point, you can query or

reset the external inhibit using the

[:SOURce]:IQ:TRIGger:INHibit

command.

Return Format

The query returns the external inhibit step, for example, 555.

Example

:IQ:TRIG:INH:STEP 555
:IQ:TRIG:INH:STEP?

Related

Command

[:SOURce]:IQ:TRIGger:INHibit

Advertising