[:SOURce]:SWEep:STEP:STOP:LEVel

Syntax

[:SOURce]:SWEep:STEP:STOP:LEVel <value>
[:SOURce]:SWEep:STEP:STOP:LEVel?

Description

Set the stop level of the sweep.
Query the stop level of the sweep.

Parameter

Name

Type

Range

Default

<value>

Real

-140dBm to 25dBm

-20dBm

Explanation

 When <value> is set in "Number" form (for example, 2), the default unit is

dBm. When <value> is set in "Number + Unit" form (for example, 2dBm), the

start level displayed in the interface of the RF signal generator is related to the

setting of Level Unit.

When the level unit is "dBm", 2.00dBm is displayed.

When the level unit is "dBmV", 48.99dBmV is displayed.

When the level unit is "dBuV", 108.99dBuV is displayed.

When the level unit is "Volts", 281.50mV is displayed.

When the level unit is "Watts", 1.58mW is displayed.

 The default unit of the return value is dBm.

 After setting the stop level, you can use the up/down direction keys or knob to

modify the stop level at the current step. At this point, you can query or set the

current step using the

[:SOURce]:SWEep:STEP:STOP:LEVel:STEP

command.

Return Format

The query returns the stop level of the sweep, for example, 2.000000.

Example

:SWE:STEP:STOP:LEV 2dBm
:SWE:STEP:STOP:LEV?

Related

Command

[:SOURce]:SWEep:STEP:STOP:LEVel:STEP